Part of the Oxford Instruments Group
Expand

Nova 96000 Series 90kV Water-Cooled Microfocus X-ray Source

The Nova Microfocus System 96000 Series is a 90kV, 80W, water-cooled X-ray source designed for applications where high power, high magnification and small spot size are important. Operated by an external high voltage Smart Controller capable of providing variable voltage and power control, the Nova Microfocus System delivers exceptional magnification and image quality with full control of “Brightness.” Maximum flux output is maintained through automatic matching of a given power setting to a corresponding optimal spot size.


    Request Product Quotes ▼

    Specifications

    Operating Voltage Range See Product Quote table below.
    Maximum Power See Product Quote table below.
    Maximum beam current 2.0mA
    Focal spot size 14-20µm @ maximum voltage and minimum power
    Focus to Object Distance (FOD) 4mm
    Cone of illumination 50° x 74° (nominal - see chart on datasheet)
    Window material and thickness Be, 254µm
    Window diameter (unobstructed) 9.5mm (0.37”)
    Window configuration 终端窗口
    Target material See Product Quote table below.
    Ambient operating temperature 10°C to 40°C
    Maximum operating temperature (anode) 70°C
    Cooling method H2O 0.15 l/min @ 15 psi
    Shielding Not shielded
    Dimensions 392.4mm L x Ø114.3mm (15.5” L x Ø4.5”)
    Weight ≤ 4kg (8.81lbs.)
    Storage conditions -10°C to 55°C, Barometric Pressure: 50-106kPa; Humidity: 10-90% (no condensation), Condensation on Be window will cause window corrosion, vacuum loss and X-ray tube failure.
    • Nova 96000系列是一个完全集成的90千伏X-ray source. Its high voltage power supply and controller provide variable control of high voltage from 10-90kV and beam current from .33-2 milliamps with full control of “Brightness”. The Smart Controller calculates spot size for a given power setting for maximum flux output.
    • Voltage and current rating (90kV, 2.0mA) are subject to maximum power dissipation rating of 80W. The X-ray tube assembly is sealed, water-cooled, and rated for continuous operation.
    • X-ray microfocus spot size is continuously adjustable from 14µm to 20µm. Power derating is provided at small spot sizes but source power is greater than or equal to 20W for a 20µm spot size.
    • The anode target material is comprised of Tungsten as standard, however a molybdenum target is also available. The target is inclined at a takeoff angle of 15 degrees with respect to the electron beam, and the exit window is aligned at an angle of 30 degrees with respect to the electron beam, so that a round microfocus X-ray spot is projected through the exit window.
    • The stability of the microfocus X-ray spot shall be less than 5µm RMS over a period of 8 hours, as verified by test. A warm-up time of up to two hours is necessary in order to meet this specification.
    • The system is supplied with a 254 micron Be exit window, allowing for close coupling (4mm) of object with the anode X-ray spot.
    • LabVIEW RT Software Interface: The Smart Controller is outfitted with a software package that provides remote control of the various functions, such as kV, mA, Brightness, power etc. It includes an RS232 Communication package and an RT version of National Instruments LabVIEW. See Software control datasheet for complete description.

    Exceptional magnification and image quality-

    High power operation- ideal for high flux applications and experiments

    Integrated package- eliminates HV cable for improved reliability

    Complete range of user control- ideal for research applications

    Compact, lightweight design- ideal for portable applications

    • Microtomography
    • Microdiffraction
    • Microfluorescence
    • CT imaging for life sciences and industrial inspection

    How to Request Product Quotes

    1. In the table below, pleaseclick the “Add to quote list” buttonfor all products that you would like quoted.

    2. Scroll to the bottom of the table andclick the green “My Quote List” buttonto visit the submission form page.

    Product Name Part Number Outline Drawing Target Operating Range (kV) Max Anode Current (mA) Max Anode Power (W) Power Density Spot Size (µm)**
    X-ray Microfocus, Nova 600, 90kV, W 96013* 8240 W 20 - 90 2.0 80 2.5W/µm 20 Max. Add to quote list
    X-ray Microfocus, Nova 600, 60kV, Mo 96016* 8240 Mo 20 - 60 2.0 60 1.5W/µm 20 Max. Add to quote list

    Note: Part number specific copies of outline drawings and product specification sheets are available.
    *Includes a thermal switch which adds an additional level of protection to the cooling system safeguards.
    **Max. = Maximum, Typ. = Typical, Nom. = Nominal (per IEC60336,NEMA XR5-1999)